Signal Conditioning
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
:INPut[n][:CHANnel[m]]:WAVelength
:INPut[n][:CHANnel[m]]:WAVelength<wsp><value>[PM | NM | UM| MM | M] | MIN |
MAX | DEF
Sets the attenuator module's operating wavelength.
This value is used to compensate for the wavelength dependence of the filter, and to calcu-
late a wavelegth dependent offset from the user offset table (if enabled).
<value>
The wavelength in meters (if you do not specify a unit).
MIN | MAX | DEF
Sets the wavelength to the module limits, or the module default.
none
INP1:WAV +1.55000000E-006
All attenuator modules.
:INPut[n][:CHANnel[m]]:WAVelength?
:INPut[n][:CHANnel[m]]:WAVelength?<wsp>MIN | MAX | DEF
Without the optional parameter, queries the operating wavelength of the attenuator.
MIN | MAX | DEF Queries the operating wavelength limits, or the module default.
4 byte Intel floating point; wavelength in m.
INP1:WAV
+1.55000000E-006<END>
All attenuator modules.
:OUTPutn[:CHANnel[m]]:APMode
:OUTPutn[:CHANnel[m]]:APMode<wsp><OFF(0) | ON(1)>
The use of this command is optional and has no effect on operation.
Included for compatibility with Agilent 8156A mainframe.
OFF or 0
ON or 1
none
OUTP1:APMode OFF
All attenuator modules.
Measurement Operations & Settings