:SENSe[n]:[CHANnel[m]]:CORRection[:LOSS][:INPut][:MAGNitude]
:SENSe[n]:[CHANnel[m]]:CORRection[:LOSS][:INPUT][:MAGNitude]<wsp>
<value>[DB|MDB]
Enters a calibration value for a module.
The calibration factor as a float value
If no unit type is specified, decibels (dB) is implied.
none
sens1:corr 10DB
All power meters
Master and slave channels are independent.
:SENSe[n]:[CHANnel[m]]:CORRection[:LOSS][:INPut][:MAGNitude]?
:SENSe[n]:[CHANnel[m]]:CORRection[:LOSS][:INPUT][:MAGNitude]?
Returns the calibration factor for a module.
none
The calibration factor as a float value. Units are in dB, although no units are returned in the
response message.
sens1:corr?
+1.00000000E+000<END>
All power meters
Master and slave channels are independent.
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO
Zeros the electrical offsets for a power meter or return loss module.
none
none
sens1:corr:coll:zero
All power meters and return loss modules
Can only be sent to master channel, slave channel is also zeroed.
Measurement Functions – The SENSe Subsystem