Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO?
:SENSe[n]:[CHANnel[m]]:CORREction:COLLect:ZERO?
Returns the status of the most recent zero command.
none
0:
any other number:
sens1:corr:coll:zero? → 0<END>
All power meters and return loss modules
Master and slave channels are independent.
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO:ALL
SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO:ALL
Zeros the electrical offsets for all installed power meter and return loss modules.
none
none
sens:chan:corr:coll:zero:all
All power meters and return loss modules
Command is independent of channel.
Setting parameters for the logging function sets some parameters,
N O T E
including hidden parameters, for the stability and MinMax functions and
vice versa. You must use the
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing command to set
parameters before you start a logging function using the
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe command.
zero succeeded without errors.
remote zeroing failed (the number is the error code returned from
the operation).
Measurement Operations & Settings
93