Agilent Technologies 8163A Programming Manual page 128

Lightwave multimeter; lightwave measurement system; lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Measurement Operations & Settings
command:
syntax:
N O T E
description:
parameters:
response:
example:
affects:
command:
syntax:
N O T E
description:
parameters:
response:
example:
affects:
128
[:SOURce[n]][:CHANnel[m]]:POWer:WAVelength
[:SOURce[n]][:CHANnel[m]:POWer:WAVelength[<wsp>
EXTernal|LOWer|UPPer|BOTH|0|1|2|3]
For compatibility reasons, WAVelength may be replaced with WAVE.
Sets the wavelength source for a dual-wavelength laser source.
EXTernal: or 0
LOWer: or 1
UPPer: or 2
BOTH: or 3
none
sour2:pow:wav upp
All dual-wavelength laser source modules and return loss modules with two internal sourc-
es
[:SOURce[n]][:CHANnel[m]]:POWer:WAVelength?
[:SOURce[n]][:CHANnel[m]:POWer:WAVelength?
For compatibility reasons, WAVelength may be replaced with WAVE.
Returns the wavelength source for a dual-wavelength laser source.
none
LOW
UPP
BOTH
sour2:pow:wav? → LOW<END>
All dual-wavelength laser source modules and return loss modules with two internal sourc-
es
Signal Generation – The SOURce Subsystem
External
The lower wavelength source
The upper wavelength source
Both wavelength sources
The lower wavelength source
The upper wavelength source
Both wavelength sources
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition