:OUTPutn[:CHANnel[m]]:STATe:APOWeron?
:OUTPutn[:CHANnel[m]]:STATe:APOWeron?
Queries the state of the shutter at power on.
none
boolean 0 The shutter is open after mainframe power on.
1 The shutter is closed after mainframe power on.
OUTP1:APOW?
0<END>
All attenuator modules.
:OUTPutn[:CHANnel[m]]:ATIMe
:OUTPutn[:CHANnel[m]]:ATIMe<wsp><value>[ NS | US | MS| S ]
Sets the powermeter averaging time, which can, if the attenuator's power contol feature is
activated, affect how the attenuator compensates for changes to input power.
<value>
The averaging time (in seconds if no unit specified).
none
OUTP1:ATIM 1s
Attenuator modules with power control.
:OUTPutn]:CHANnel[m]]:ATIMe?
:OUTPutn[:CHANnel[m]]:ATIMe?
Queries the powermeter averaging time.
none
4 byte Intel floating point; the averaging time in seconds
OUTP1:ATIM?
1<END>
Attenuator modules with power control.
:OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO
:OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO
Zeros the electrical offsets of the attenuator's integrated powermeter.
none
none
OUTP1:CORR:COLL:ZERO