Measurement Functions – The SENSe Subsystem
command:
syntax:
description:
parameters:
N O T E
N O T E
N O T E
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
Setting parameters for the MinMax function sets some parameters,
N O T E
including hidden parameters, for the stability and logging functions and
vice versa. You must use the
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:MINMax command to set
parameters before you start a MinMax function using the
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe command.
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:MINMax
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:MINMax<wsp>
CONTinous|WINDow|REFResh,<data points>
Sets the MinMax mode and the number of data points for the
MinMax data acquisition function.
CONTinous:
WINDow:
REFResh:
Data Points is the number of samples that are recorded in the memory buffer used by the
WINDow and REFResh modes. Data Points is an integer value.
See Chapter 3 of the Agilent 8163A/B Lightwave Multimeter, Agilent 8164A/B Lightwave
Measurement System, & Agilent 8166A/B Lightwave Multichannel System User's Guide, for
more information on MinMax mode.
See ":SENSe[n][:CHANnel[m]]:FUNCtion:STATe" on page 100 for information on start-
ing/stopping a data acquisition function.
See ":SENSe[n][:CHANnel[m]]:FUNCtion:RESult?" on page 97 for information on accessing
the results of a data acquisition function.
See "Triggering and Power Measurements" on page 171 for information on how triggering
affects data acquisition functions.
none
sens1:func:par:minm WIND,10
All power meters and return loss modules
Can only be sent to master channel, slave channel is also affected.
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:MINMax?
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:MINMax?
Returns the MinMax mode and the number of data points for the MinMax data acquisition
function.
none
CONT:
WIND:
REFR:
The number of data points is returned as an integer value.
Measurement Operations & Settings
continuous MinMax mode
window MinMax mode
refresh MinMax mode
continuous MinMax mode
window MinMax mode
refresh MinMax mode
95