Table 9 Generating Output Triggers From Power Measurements - Agilent Technologies 8163A Programming Manual

Lightwave multimeter; lightwave measurement system; lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Measurement Operations & Settings

Table 9 Generating Output Triggers from Power Measurements

Hardware
Trigger Rearming
Triggering
trig:outp
trig:outp:rearm
DISabled
-
AVGover
ON
MEASure
ON
AVGover
OFF
MEASure
OFF
command:
syntax:
description:
parameters:
N O T E
N O T E
response:
example:
172
Software Triggering
init:imm
An output trigger will never be generated.
An output trigger is generated for every new power measurement when the aver-
aging time period finishes.
An output trigger is generated for every new power measurement when the aver-
aging time period begins.
An output trigger is generated when the averaging
time period of the first power measurement finish-
es. A further hardware output trigger cannot be
generated until you send trig:outp:rearm.
An output trigger is generated when the averaging
time period of the first power measurement begins.
A further hardware output trigger cannot be gener-
ated until you send trig:outp:rearm.
:TRIGger
:TRIGger<wsp>NODEA|1|NODEB|2
Generates a hardware trigger.
1 or NODEA:
Is identical to a trigger at the Input Trigger Connector.
2 or NODEB:
Generates trigger at the Output Trigger Connector.
A hardware trigger cannot be effective in the DISabled triggering mode but can be
effective in DEFault, PASSthrough or LOOPback triggering modes, see
":TRIGger:CONFiguration" on page 177 for information on triggering modes.
":TRIGger" on page 179 describes the :TRIGger command for advanced users using
":TRIGger:CONFiguration:EXTended" on page 179.
none
trig 1
Data Acquisition Functions
sens:func:stat
init:cont
MINMax
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
Triggering - The TRIGger Subsystem
LOGGing|STABility
Applies for all subsequent
data acquisition functions.
Applies for all subsequent
data
acquisition functions.
An output trigger is generat-
ed for every new power mea-
surement when the
averaging time period finish-
es. Applies for all subsequent
data acquisition functions.
An output trigger is generat-
ed for every new power mea-
surement when the
averaging time period begins.
Applies for all subsequent
data acquisition functions.