[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:[STATe]?
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:[STATe]?
Returns the state of a sweep.
none
+0:
+1:
wav:swe?
+0<END>
All tunable laser modules
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:NEXT
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:NEXT
Performs the next sweep step in stepped sweep if it is paused or in manual sweep
none
none
wav:swe:step:next
All tunable laser modules
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:PREVious
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:PREVious
Performs one sweep step backwards in stepped sweep if it is paused or in manual sweep
none
none
wav:swe:step:prev
All tunable laser modules
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:[WIDTh]
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:[WIDTh]<wsp>
<value>[PM|NM|UM|MM|M]
Sets the width of the sweep step.
In continuous sweep mode, the end of a step is used for triggering.
The width of the sweep step as a float value.
If you specify no units in your command, meters are used as the default.
none
wav:swe:step 5nm
All tunable laser modules
Signal Generation – The SOURce Subsystem
Sweep is not running
Sweep is running