Agilent Technologies 8163A Programming Manual page 72

Lightwave multimeter; lightwave measurement system; lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Instrument Setup and Status
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
72
:STATus:QUEStionable:ENABle[:LEVel0]
:STATus:QUEStionable:ENABle[:LEVel0]<wsp><value>
Sets the bits in the Questionable Status Enable Summary Mask (QSESM) that enable the
contents of the QSESR to affect the Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the QSESR to affect bit 3 of
the Status Byte.
The bit value for the questionable enable mask as a 16-bit signed integer value (0 .. +32767)
The default value is 0.
none
stat:ques:enab 128
:STATus:QUEStionable:ENABle[:LEVel0]?
:STATus:QUEStionable:ENABle[:LEVel0]?
Returns the QSESM for the event register
none
The bit value for the QSEM as a 16-bit signed integer value (0 .. +32767)
stat:ques:enab? → +128<END>
:STATus:QUEStionable[:EVENt]:LEVel1?
:STATus:QUEStionable[:EVENt]:LEVel1?
Returns the Questionable Status Event Summary Register (QSESR) for slots 15 to 17 of the
Agilent 8166A/B Lightwave Multichannel System.
none
The sum of the results for the slots (a 16-bit signed integer value, where 0 ≤ value ≤ 32767):
Bits
Mnemonics
Agilent 8166A/B
15-4
Not used
3
Slot 17 Summary
2
Slot 16 Summary
1
Slot 15 Summary
0
Not used
stat:ques:level1? → +0<END>
Status Reporting – The STATus Subsystem
Decimal Value
0
8
4
2
0