Measurement Operations & Settings
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
152
:OUTPutn[:CHANnel[m]]:APMode?
:OUTPutn[:CHANnel[m]]:APMode?
Queries whether the user has amended the power value or the attenuation value.
This use of this command is optional.
Included for compatibility with Agilent 8156A mainframe.
none
boolean 0 User has amended the attenuation value.
1 User has amended the power value.
OUTP1:APMode?
0<END>
All attenuator modules.
:OUTPutn[:CHANnel[m]]:POWer
:OUTPutn[:CHANnel[m]]:POWer<wsp><value>[PW | NW | UW | MW | W | DBM ] | MIN | MAX |
DEF
Sets the output power value (P ).
If your attenuator module does not include the power control feature, the new filter attenua-
α
tion (
) is calculated from the reference power (P
filter (new)
P
(dBm) = P
(dBm) -
set(new)
ref
If your attenuator module includes the power control feature, the filter attenuation is changed
until the set power (measured by the module's internal power meter) has been reached.
P
(dBm) = P
set(new)
att(new)
If the set power cannot be achieved ExP (indicating 'Excessive Power' ) is displayed, and the
appropriate GPIB status bit is set. The status of these bits can be queried using ":STA-
Tus:QUEStionable:CONDition[:LEVel0]?" on page 71
<value>
Desired output power (if unit not specified current unit is used).
MIN | MAX | DEF
Sets the output power to the module limits, or the module default.
none
OUTP1:POW 12
All attenuator modules.
:OUTPutn[:CHANnel[m]]:POWer?
:OUTPutn[:CHANnel[m]]:POWer<wsp>MIN | MAX | DEF
Without the optional parameter, queries the output power value.
MIN | MAX | DEF Queries the output power module limits, or the module default.
4 byte Intel floating point; output power in current power unit.
OUTP1:POW?
12<END>
All attenuator modules.
α
(dB) - P
filter (new)
Offset
(dBm) - P
(dB)
offset
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
Signal Conditioning
) :
ref
(dB)

Hide quick links: