110
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
For the currently selected source, start the calibration and save the calibration values for a
defined termination reference measurement. See
":SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]" on page 112 for infor-
mation on setting the return loss value of your reference reflector.
none
none
sens1:ret:cal:coll:term
All return loss modules
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:TERMination?
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:TERMination?
Queries the T-value (termination calibration value) for the return loss module
none
Termination calibration value as a float in dB
sens1:ret:cal:term?
+6.5000E+001
All return loss modules
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:VALues?
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:VALues?
Returns the the current calibration values
1. monitor diode reference power
2. return loss diode reference power
3. monitor diode parasitics power
4. return loss diode parasitics power.
Returns power values in W
none
sens1:ret:cal:val
All return loss modules
Measurement Functions – The SENSe Subsystem
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition