Agilent Technologies 8163A Programming Manual page 100

Lightwave multimeter; lightwave measurement system; lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Measurement Operations & Settings
command:
syntax:
description:
parameters:
N O T E
N O T E
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
100
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe<wsp>
LOGGing|STABility|MINMax,STOP|STARt
Enables/Disables the logging, MinMax, or stability data acquisition function mode.
LOGGing:
STABility:
MINMax:
STOP:
STARt:
When you enable a logging data acquisition function for a Agilent 8163A/B Series Power
Meter with averaging time of less than 100 ms with input hardware triggering disabled, all
GPIB commands will be ignored for the duration of the function.
See ":SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing" on page 94 for more infor-
mation on the logging data acquisition function.
Stop any function before you try to set up a new function. Some parameters cannot be set
until you stop the function.
none
sens1:func:stat logg,star
All power meters and return loss modules
Can only be sent to master channel, slave channel is also affected.
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe?
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe?
Returns the function mode and the status of the data acquisition function.
none
NONE
LOGGING_STABILITY
MINMAX
PROGRESS
COMPLETE
sens1:func:stat?
LOGGING_STABILITY,COMPLETE<END
All power meters and return loss modules
Can only be sent to master channel, slave channel parameters are identical.
Measurement Functions – The SENSe Subsystem
Logging data acquisition function
Stability data acquisition function
MinMax data acquisition function
Stop data acquisition function
Start data acquisition function
No function mode selected
Logging or stability data acquisition function
MinMax data acquisition function
Data acquisition function is in progress
Data acquisition function is complete
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
>