Measurement Operations & Settings
command:
syntax:
description:
response:
response
format:
example:
affects:
command:
syntax:
description:
response
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
84
:SLOT[n]:HEAD[m]:WAVelength:RESPonse:CSV?
:SLOT[n]:HEAD[m]:WAVelength:RESPonse:CSV?
Returns the wavelength response from the attenuator module in CSV format.
Wavelength Response table as a string
The string is a comma separated value (CSV) list and can be written to a file and be pro-
cessed with a spreadsheet program.
List format:
λ1, c1\n
λ2, c2\n
.......
λn, cn\n
"," separates wavelength and response factor
"\n" = ASCII code 10 separate value pairs
slot1:head1:wav:resp:csv?
1.956\n...
Attenuator with power control, all powermeters, return loss modules
:SLOT[n]:HEAD[m]:WAVelength:RESPonse:SIZE?
:SLOT[n]:HEAD[m]:WAVelength:RESPonse:SIZE?
Returns the number of elements in the wavelength response table.
Number of elements in the wavelength table as an integer value
slot2:head1:wav:resp:size?
Attenuator with power control, all powermeters, return loss modules
:SPECial:REBoot
:SPECial:REBoot
Reboots the mainframe and all modules.
none
none
spec:reb
1200e-6,2.019\n 1210e-6,
50<END>