Measurement Operations & Settings
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
162
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:REFerence?
:CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:REFerence?
Queries the currently selected slot and channel of the external powermeter (hosted in the
same mainframe as the attenuator module) used by TOREF.
See: ":CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:VALue" on page 161
none
the slot and channel of the external powermeter as integer values.
CONF1:OFFS:WAV:REF?
Attenuator modules with power control.
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:VALue:WAVelength?
:CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:VALue:WAVelength?<wsp><index>
Queries a wavelength value from its position, or index, in the offset table. Offset table entries
are ordered from shortest to longest wavelength. The first index number = 1.
<index>
The position of the wavelength value in the offset table.
4 byte Intel floating point; the wavelength in meters
CONF1:OFFS:WAV:VAL:WAV? 1
All attenuator modules.
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:VALue:OFFSet?
:CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:VALue:OFFSet?<wsp><index | wavelength [PM |
NM | UM | MM| M],>
Queries an offset value from the position, or index, of the associated wavelength in the offset
table. Offset table entries are ordered from shortest to longest wavelength. The first index
number =1.
Or: Queries the offset applied for a particular wavelength.
<index>
The position of the value pair (wavelength; offset) in the offset table.
<wavelength>
The wavelength for the offset table entry, in m
4 byte Intel floating point; the offset
CONF1:OFFS:WAV:VAL:OFFS? 1
All attenuator modules.
+2,+1<END>
+1.55000000E-006<END>
2