GPIB Command Compatibility List
Table 16 Error Handling Changes
Table 17 Specific Errors
254
Error Handling
Most error commands and error texts for all instruments are new.
The HP 8153A timed out for every error. Errors are handled differently by
the Agilent 8163A/B and 8164A/B; instead of timing out for every error,
special values are returned for erroneous queries. Table 16 and Table 17
detail the new errors
The error queue is written to as before.
Expected Return Value
FLOAT(32/64)
(U)INT(16/32)
Block
Boolean Value
Enum
Command
FETCh:POWer? - without
using a preceding trigger
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
Returned Value
FLT/DBL_MAX
(U)INT(16/32)_MAX
" "
0
Time out
Change
Returns the last valid value instead of timing
out.
No error is generated.
Error Handling
Affects
8153
8153
8153
8153
8153
Affects
8153