Signal Generation – The SOURce Subsystem
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:AUTO
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:AUTO<wsp>OFF|ON|0|1
Selects Automatic or Manual Attenuation Mode.
In Automatic Attenuation Mode, you specify the output power.
In Manual Attenuation Mode, you must specify both the laser output power, and the attenu-
ation level.
OFF or 0:
ON or 1:
none
sour0:pow:att:auto 1
All tunable laser sources with a built-in optical attenuator.
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:AUTO?
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:AUTO?
Queries whether the instrument is in Automatic or Manual Attenuation Mode.
none
0:
Manual Attenuation Mode
1:
Automatic Attenuation Mode
→
sour0:pow:att:auto?
1<END>
All tunable laser modules with a built-in optical attenuator.
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:DARK
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:DARK<wsp>OFF|ON|0|1
Sets or unsets the attenuator to 'dark' position.
Dark position blocks all light from the laser. You can use this as an alternative to disabling
the laser, the advantage of doing this is that you avoid the laser rise time.
This command is available in Attenuation Mode Only.
OFF or 0:
ON or 1:
none
sour0:pow:att:dark 1
All tunable laser modules with a built-in optical attenuator.
Attenuation Mode
Power Mode
Unsets dark position
Sets dark position
Measurement Operations & Settings