Agilent Technologies 8163A Programming Manual page 123

Lightwave multimeter; lightwave measurement system; lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Signal Generation – The SOURce Subsystem
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:AUTO
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:AUTO<wsp>OFF|ON|0|1
Selects Automatic or Manual Attenuation Mode.
In Automatic Attenuation Mode, you specify the output power.
In Manual Attenuation Mode, you must specify both the laser output power, and the attenu-
ation level.
OFF or 0:
ON or 1:
none
sour0:pow:att:auto 1
All tunable laser sources with a built-in optical attenuator.
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:AUTO?
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:AUTO?
Queries whether the instrument is in Automatic or Manual Attenuation Mode.
none
0:
Manual Attenuation Mode
1:
Automatic Attenuation Mode
sour0:pow:att:auto?
1<END>
All tunable laser modules with a built-in optical attenuator.
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:DARK
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]:DARK<wsp>OFF|ON|0|1
Sets or unsets the attenuator to 'dark' position.
Dark position blocks all light from the laser. You can use this as an alternative to disabling
the laser, the advantage of doing this is that you avoid the laser rise time.
This command is available in Attenuation Mode Only.
OFF or 0:
ON or 1:
none
sour0:pow:att:dark 1
All tunable laser modules with a built-in optical attenuator.
Attenuation Mode
Power Mode
Unsets dark position
Sets dark position
Measurement Operations & Settings
123

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8166a8166b8163b8164a8164b

Table of Contents