Agilent Technologies 8163A Programming Manual page 65

Lightwave multimeter; lightwave measurement system; lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Status Reporting – The STATus Subsystem
command:
syntax:
description:
parameters:
response:
example:
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
:STATus:OPERation:CONDition[:LEVel0]?
:STATus:OPERation:CONDition[:LEVel0]?
Reads the Operational Status Condition Summary Register.
none
The sum of the results for the individual slots (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bits
Mnemonics
Agilent 8163A/B
15
Not used
14
Not used
13
Not used
12
Not used
11
Not used
10
Not used
9
Not used
8
Not used
7
Not used
6
Not used
5
Not used
4
Not used
3
Not used
2
Slot 2 Summary
1
Slot 1 Summary
0
Not used
stat:oper:cond? → +0<END>
Agilent 8164A/B
Agilent 8166A/B
Not used
Not used
Not used
Slot 14 Summary
Not used
Slot 13 Summary
Not used
Slot 12 Summary
Not used
Slot 11 Summary
Not used
Slot 10 Summary
Not used
Slot 9 Summary
Not used
Slot 8 Summary
Not used
Slot 7 Summary
Not used
Slot 6 Summary
Not used
Slot 5 Summary
Slot 4 Summary
Slot 4 Summary
Slot 3 Summary
Slot 3 Summary
Slot 2 Summary
Slot 2 Summary
Slot 1 Summary
Slot 1 Summary
Slot 0 Summary
Level 1 Summary
Instrument Setup and Status
Decimal Value
0
16384
8192
4096
2048
1024
512
256
128
64
32
16
8
4
2
1
65

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8166a8166b8163b8164a8164b

Table of Contents