Agilent Technologies 8163A Programming Manual page 163

Lightwave multimeter; lightwave measurement system; lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Signal Conditioning
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
N O T E
parameters:
response:
example:
affects:
command:
syntax:
CAUTION
description:
parameters:
response:
example:
affects:
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:VALue:PAIR?
:CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:VALue:PAIR?<wsp><index | wavelength[PM | NM |
UM | MM| M],>
Queries an offset value pair (wavelength:offset) from the position, or index, of the associated
wavelength in the offset table. Offset table entries are ordered from shortest to longest wave-
length.
Or: Queries the offset value pair (wavelength:offset) applied for the specified wavelength.
<index>
The position of the wavelength; offset value pair in the offset table.
<wavelength>
The wavelength for the offset table entry, in m
char$ in SCPI block format (Intel byte order); wavelength:offset
CONF1:OFFS:WAV:VAL:PAIR? 1
All attenuator modules.
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:VALue:DELete
:CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:VALue:DELete<wsp><index | wavelength[PM | NM
| UM | MM| M],>
Deletes an offset value pair (wavelength:offset) from the position, or index, of the associated
wavelength in the offset table. Offset table entries are ordered from shortest to longest wave-
length.
Or: Deletes the offset value pair (wavelength:offset) associated with the specified wavelength.
Deleting a value pair decrements the index value of every subsequent value pair by 1. When us-
ing this command, you may prefer to work from large to small index values.
<index>
The position of the wavelength:offset value pair in the offset table.
<wavelength>
The wavelength for the offset table entry, in m
none
CONF1:OFFS:WAV:VAL:DEL 1
All attenuator modules.
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:VALue:DELete:ALL
:CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:VALue:DELete:ALL
This command clears the offset table!
Deletes every value pair (wavelength:offset) from the offset table.
none
none
CONF1:OFFS:WAV:VAL:DEL:ALL
All attenuator modules.
"+1.55000000E-006:2"
Measurement Operations & Settings
163