Agilent Technologies 8163A Programming Manual page 120

Lightwave multimeter; lightwave measurement system; lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Measurement Operations & Settings
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
120
[:SOURce[n]][:CHANnel[m]]:FM:SBSCtrl:FREQuency[l]
[:SOURce[n]][:CHANnel[m]]:FM:SBSCtrl:FREQuency[l]<wsp><frequency>
[MHZ|KHZ|HZ|MIN|MAX|DEF]
Sets the frequency of the SBS Control modulation.
Enable frequency modulation before issuing this command.
Refer to "[:SOURce[n]][:CHANnel[m]]:FM:SOURce[l]" on page 118 and
"[:SOURce[n]][:CHANnel[m]]:FM:STATe[l]" on page 119.
The modulation frequency as a float value.
The default units are HZ, although KHZ, MHZ, GHZ and THZ can also be specified.
Also allowed are:
none
sour2:fm:sbsc:freq 4000Hz
Agilent 81980A, 81940A, 81989A, 81949A compact tunable lasers.
[:SOURce[n]][:CHANnel[m]]:FM:SBSCtrl:FREQuency[l]?
[:SOURce[n]][:CHANnel[m]]:FM:SBSCtrl:FREQuency[l]?<wsp>[MIN|MAX|DEF]
Queries the currently set frequency of the SBS Control modulation.
Optional
The modulation frequency in Hz as a float value
sour2:fm:freq?
+4.00000E+03<END>
Agilent 81980A, 81940A, 81989A, 81949A compact tunable lasers.
[:SOURce[n]][:CHANnel[m]]:FM:SBSCtrl:LEVel[l]
[:SOURce[n]][:CHANnel[m]]:FM:SBSCtrl:LEVel[l]<wsp>[MIN|MAX|DEF]
Sets the excursion of the SBS Control frequency modulation to a percentage of its maximum
value.
The excursion level as a percentage of its maximum value.
Also allowed:
none
sour2:fm:sbsc:lev 80
Agilent 81980A, 81940A, 81989A, 81949A compact tunable lasers.
Signal Generation – The SOURce Subsystem
MIN: minimum programmable value
MAX: maximum programmable value
DEF: default preset (*RST) value.
MIN: returns the minimum programmable value
MAX: returns the maximum programmable value
DEF: returns the default preset (*RST) value.
MIN: minimum programmable value (0%)
MAX: maximum programmable value (100%)
DEF: default preset (*RST) value.