Agilent Technologies 8163A Programming Manual page 111

Lightwave multimeter; lightwave measurement system; lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Measurement Functions – The SENSe Subsystem
command:
syntax:
description:
N O T E
parameters:
response:
example
affects:
command:
syntax:
description:
N O T E
parameters:
response:
example
affects:
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l]
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l]<wsp><value>[dB]
Sets the front panel delta, that is, the loss correction value, for example, due to the front pan-
el connector. Twice this value is added to the measured Return Loss.
Use [l] to set the front panel delta for an external source or the upper or lower wavelength la-
ser source of a dual return loss module.
An external laser source is denoted by 0. 0 is the default value of [l].
A lower wavelength source is denoted by 1.
An upper wavelength source is denoted by 2.
Sets the front panel delta as a float value in dB
none
sens1:ret:cal:corr:fpd 0.08DB
All return loss modules
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l]?
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l]?
Returns the front panel delta, that is, the loss correction value, for example, due to the front
panel connector. Twice this value is added to the measured Return Loss.
Use [l] to query the front panel delta for an external source or the upper or lower wavelength
laser source of a dual return loss module.
An external laser source is denoted by 0. 0 is the default value.
A lower wavelength source is denoted by 1.
An upper wavelength source is denoted by 2.
Returns the front panel delta as a float value in dB
none
sens1:ret:cal:corr:fpd?
All return loss modules
+8.00000000E-002<END>