Display Operations – The DISPlay Subsystem
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
N O T E
response:
example:
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
:DISPlay:BRIGhtness?
:DISPlay:BRIGhtness?
Queries the brightness of the display.
none
An integer value in the range 0 to 100
disp:brig? → +75<END>
Agilent 8163B Lightwave Multimeter and 8166B Lightwave Multichannel System -
8164A Lightwave Measurement System: only checks if the value equals 0. (0 -> display off,
other values: display on)
:DISPlay:ENABle
:DISPlay:ENABle<wsp>ON|OFF|1|0
Enables or disables the display.
The display is cleared, and an appropriate message displayed. This setting may improve
sweep performance.
A boolean value:
If you press [L
] softkey, the display is enabled automatically.
OCAL
none
disp:enab 1
:DISPlay:ENABle?
:DISPlay:ENABle?
Queries the state of the display.
none
A boolean value:
disp:enab? → 1<END>
:DISPlay:LOCKout
:DISPlay:LOCKout<wsp>ON|OFF|1|0
Enables or Disables local operation.
A boolean value:
none
disp:lock 1<END>
Mass Storage, Display, and Print Functions
OFF or boolean 0 – switch off the display
ON or boolean 1 – switch on the display
0 – the display is turned off
1 – the display is turned on
OFF or boolean 0 – local operation is disabled
ON or boolean 1 – local operation is enabled.