Signal Generation – The SOURce Subsystem
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO
Executes a wavelength zero.
none
none
sour2:wav:corr:zero
All tunable laser modules except Agilent 81649A and Agilent 81689A/B
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO:ALL
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO:ALL
Executes a wavelength zero on every tunable laser source in a mainframe.
none
none
sour2:wav:corr:zero:all
All tunable laser modules except Agilent 81649A and Agilent 81689A/B
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO:TEMPerature:ACTual?
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO:TEMPerature:ACTual?
Reports the current lambda zero tempearture.
none
float value; temperature in °C
sour0:wav:corr:zero:temp:act?
All tunable laser modules except Agilent 81649A, Agilent 81689A/B and 819xxA/B
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO:TEMPerature:DIFFerence?
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO:TEMPerature:Difference?
Reports the temperature difference required to trigger an auto lamda zero.
none
float value; temperature in °C
sour0:wav:corr:zero:temp:diff?
All tunable laser modules except Agilent 81649A, Agilent 81689A/B and 819xxA/B