Measurement Operations & Settings
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
156
:OUTPutn[:CHANnel[m]]:POWer:UNit?
:OUTPutn[:CHANnel[m]]:POWer:UNit?
Queries whether the power unit is dBm or W
none
boolean 0 The power unit is dBm
1 The power unit is W.
OUTP1:POW:UN?
0<END>
All attenuator modules.
:OUTPutn[:CHANnel[m]]:[STATe]
:OUTPutn[:CHANnel[m]]:[STATe]<wsp>OFF(0) | ON(1)
Sets the state of the shutter.
OFF or 0
Shutter closed.
ON or 1
Shutter open.
none
OUTP1:STAT OFF
All attenuator modules.
:OUTPutn[:CHANnel[m]]:[STATe]?
:OUTPutn[:CHANnel[m]]:[STATe]?
Queries the state of the shutter.
none
boolean 0 The shutter is open.
1 The shutter is closed.
OUTP1:STAT?
0<END>
All attenuator modules.
:OUTPutn[:CHANnel[m]]:STATe:APOWeron
:OUTPutn[:CHANnel[m]]:STATe:APOWeron<wsp>OFF(0) | ON(1)
Sets the state of the shutter when the mainframe is turned on.
OFF or 0
Shutter closed after mainframe power on.
ON or 1
Shutter open after mainframe power on.
none
OUTP1:APOW OFF
All attenuator modules.