[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:DWELl
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:DWELl<wsp>
<value>|MIN|MAX|DEF[NS|US|MS|S]
Sets the dwell time. Can only be used when sweep is stepped.
Cannot be set while a sweep is running.
The dwell time as a float value.
If you specify no units in your command, seconds are used as the default.
Also allowed are: MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of, the
minimum programmable value and the maximum programmable value
none
wav:swe:dwel 500ms
All tunable laser modules
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:DWELl?
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:DWELl?[<wsp>MIN|MAX|DEF]
Returns the dwell time. Can only be used when sweep is stepped.
none
Also allowed are: MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of, the
minimum programmable value and the maximum programmable value
The dwell time in seconds.
wav:swe:dwel?
+5.00000000E-001<END>
All tunable laser modules
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:EXPectedtriggers?
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:EXPectedtriggers?
Returns the number of triggers. A tunable laser wavelength sweep causes a number of trig-
gers, this number is required to configure a triggering data acquisition function on a power
meter. The number returned by this function can be used to configure a Power Meter for co-
ordinated measurements with a tunable laser source (see command
":SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing" on page 94).
none
the number of expected triggers as an unsigne integer value.
sour0:wav:swe:exp?
12001