Triggering - The TRIGger Subsystem
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
:TRIGger[n][:CHANnel[m]]:INPut:REARm?
:TRIGger[n][:CHANnel[m]]:INPut:REARm?
Returns the arming response of a channel to an incoming trigger.
none
A boolean value:
trig1:inp:rearm?
0<END>
All Agilent 8163A/B series power meter modules, and Agilent 8161x series return loss mod-
ules.
Can only be sent to master channel, slave channel parameters are identical.
:TRIGger[n][:CHANnel[m]]:OFFSet
:TRIGger[n][:CHANnel[m]]:OFFSet <value>
Sets the number of incoming triggers received before data logging begins.
<value> - an integer value. (maximum possible value is 1e+9)
none
trig1:offs 5
All Agilent 81636B and 81637B series power meter modules.
:TRIGger[n][:CHANnel[m]]:OFFSet?
:TRIGger[n][:CHANnel[m]]:OFFSet?
Returns the number of incoming triggers received before data logging begins.
none
an integer value.
trig1:offs?
5<END>
All Agilent 81636B and 81637B series power meter modules.
0: trigger rearming disabled
1: trigger rearming enabled (default)