Mass Storage, Display, and Print Functions
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
186
Display Operations – The
DISPlay Subsystem
The DISPlay subsystem lets you control what you see on the instrument's
display.
:DISPlay:CONTrast
:DISPlay:CONTrast<wsp><value>
Controls the contrast of the display.
An integer value in the range 0 to 100
none
disp:cont 50
Agilent 8163B Lightwave Multimeter and 8166B Lightwave Multichannel System
:DISPlay:CONTrast?
:DISPlay:CONTrast?
Queries the contrast of the display.
none
An integer value in the range 0 to 100
disp:cont? → +50<END>
Agilent 8163B Lightwave Multimeter and 8166B Lightwave Multichannel System
:DISPlay:BRIGhtness
:DISPlay:BRIGhtness<wsp><value>
Controls the brightness of the display.
An integer value in the range 0 to 100
none
disp:brig 75
Agilent 8163B Lightwave Multimeter and 8166B Lightwave Multichannel System -
8164A Lightwave Measurement System: only checks if the value equals 0. (0 -> display off,
other values: display on)