Status Reporting – The STATus Subsystem
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
:STATus:OPERation:CONDition:LEVel1?
:STATus:OPERation:CONDition:LEVel1?
Returns the Operational Status Condition Summary Register for slots 15 to 17 of the
Agilent 8166B Lightwave Multichannel System.
none
The sum of the results for slots 15 to 17 (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bits
Mnemonics
Agilent 8166A/B
15-4
Not used
3
Slot 17 Summary
2
Slot 16 Summary
1
Slot 15 Summary
0
Not used
stat:oper:cond:level1? → +0<END>
:STATus:OPERation:ENABle:LEVel1
:STATus:OPERation:ENABle:LEVel1<wsp><value>
Sets the bits in the Operational Status Enable Summary Mask (OSESM) that enable the con-
tents of the OSESR for slots 15 - 17 of the Agilent 8166A/B Lightwave Measurement System
to affect the Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the OSESR for slots 15 - 17 of
the Agilent 8166A/B Lightwave Measurement System to affect bit 7 of the Status Byte.
The bit value for the OSESM as a 16-bit signed integer value (0 .. +32767)
The default value is 0.
none
stat:oper:enab:level1 128
:STATus:OPERation:ENABle:LEVel1?
:STATus:OPERation:ENABle:LEVel1?
Returns the OSESM for the OSESR for slots 15 - 17 of the Agilent 8166A/B Lightwave Mea-
surement System
none
The bit value for the operation enable mask as a 16-bit signed integer value (0 .. +32767)
stat:oper:enab:level1? → +128<END>
Decimal Value
0
8
4
2
0
Instrument Setup and Status
67