Instrument Setup and Status
command:
syntax:
description:
parameters:
response:
example:
62
*TST?
*TST?
The self-TeST query *TST? makes the instrument perform a self-test and place the results of
the test in the output queue. If the self-test fails, the results are also put in the error queue.
We recommend that you read self-test results from the error queue. No further commands
are allowed while the test is running. After the self-test the instrument is returned to the set-
ting that was active at the time the self-test query was processed. The self-test does not re-
quire operator interaction beyond sending the *TST? query.
none
The sum of the results for the individual tests (a 32-bit signed integer value, where 0 ≤ value
≤ 4294967296):
Bits
Mnemonic
31
Selftest failed on Mainframe
18 - 30
Not used
17
Selftest failed on Slot 17
16
Selftest failed on Slot 16
15
Selftest failed on Slot 15
14
Selftest failed on Slot 14
13
Selftest failed on Slot 13
12
Selftest failed on Slot 12
11
Selftest failed on Slot 11
10
Selftest failed on Slot 10
9
Selftest failed on Slot 9
8
Selftest failed on Slot 8
7
Selftest failed on Slot 7
6
Selftest failed on Slot 6
5
Selftest failed on Slot 5
4
Selftest failed on Slot 4
3
Selftest failed on Slot 3
2
Selftest failed on Slot 2
1
Selftest failed on Slot 1
0
Selftest failed on Slot 0
If 16 is returned, the module in slot 4 has failed.
If 18 is returned, the modules in slots 1 and 4 have failed.
A value of zero indicates no errors.
*TST? → 0<END>
Decimal Value
A negative value
0
131072
65536
32768
16384
8192
4096
2048
1024
512
256
128
64
32
16
8
4
2
1