Triggering - The TRIGger Subsystem
command:
syntax:
description:
N O T E
parameters:
N O T E
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example:
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
:TRIGger[n][:CHANnel[m]]:OUTPut:REARm
:TRIGger[n][:CHANnel[m]]:OUTPut:REARm<wsp>OFF|ON|0|1
Sets the arming response of a channel to an outgoing trigger.
See Table Table 9 , for information on how this command affects the generation of
output triggers using power measurements.
A boolean value:
If you return to Local control, all modules return to the default setting.
none
trig1:outp:rearm 1
All Agilent 8163A/B series power meters, and Agilent 8161x series return loss modules.
Can only be sent to master channel, slave channel is also affected.
:TRIGger[n][:CHANnel[m]]:OUTPut:REARm?
:TRIGger[n][:CHANnel[m]]:OUTPut:REARm?
Returns the arming response of a channel to an outgoing trigger.
none
A boolean value:
trig1:outp:rearm?
0<END>
All Agilent 8163A/B series power meters, and Agilent 8161x series return loss modules.
Can only be sent to master channel, slave channel parameters are identical.
:TRIGger:CONFiguration
:TRIGger:CONFiguration<wsp><triggering mode>
Sets the hardware trigger configuration with regard to Output and Input Trigger Connectors.
0 or DISabled:
Trigger connectors are disabled.
1 or DEFault:
The Input Trigger Connector is activated, the incoming trigger response
for each slot ":TRIGger[n][:CHANnel[m]]:INPut" on page 173 deter-
mines how each slot responds to an incoming trigger, all slot events
(see ":TRIGger[n][:CHANnel[m]]:OUTPut" on page 176) can trigger the
Output Trigger Connector.
2 or PASSthrough:
The same as DEFault but a trigger at the Input Trigger Connector gener-
ates a trigger at the Output Trigger Connector automatically.
3 or LOOPback:
The same as DEFault but a trigger at the Output Trigger Connector gen-
erates a trigger at the Input Trigger Connector automatically.
none
trig:conf dis
OFF or 0: trigger rearming disabled
ON or 1: trigger rearming enabled (default)
0: trigger rearming disabled (default)
1: trigger rearming enabled.
Measurement Operations & Settings