Illegal Temperature Value from Sensor at Air Inlet

Error ID: 1072
This temperature sensor delivered a value outside the allowed range. The
parameter of this event equals the measured temperature in 1/100
centigrade. As a result the temperature control is switched off.
Probable cause
1
2

Heater at fan assembly failed

Error ID: 1073
Every time the deuterium lamp or the tungsten lamp (DAD only) is
switched on or off a heater self- test is performed. If the test fails an error
event is created. As a result the temperature control is switched off.
Probable cause
1
2
Agilent 1200 Infinity Series VWD - User Manual
The temperature sensor is defect.
Detector is exposed to illegal ambient
conditions.
Defective connector or cable.
Defective heater.
Error Information
Detector Error Messages
Suggested actions
• Replace the cable to the main board.
• Please contact your Agilent service
representative.
Verify that the ambient conditions are within
the allowed range.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.