Download Print this page

Thermo Scientific 42i Instruction Manual page 281

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Thermo Fisher Scientific
Send:
gas mode
Receive:
gas mode sample
set sample
set sample gas
These commands set the zero/span valves to the sample mode. The
example below sets the instrument to sample mode, that is, the instrument
is reading the sample gas.
Send:
set sample
Receive:
set sample ok
set zero
set zero gas
These commands set the zero/span valves to the zero mode. The example
below sets the instrument to zero mode that is, the instrument is reading
the sample gas.
Send:
set zero
Receive:
set zero ok
set span
set span gas
These commands set the zero/span valves to the span mode. The example
below sets the instrument to span mode that is, the instrument is sampling
span gas.
Send:
set span
Receive:
set span ok
gas unit
This command reports the current gas units (ppb or μg/m
reports that the gas unit is set to ppb.
Send:
gas unit
Receive:
gas unit ppb
set gas unit
3
unit = | ppb | μg/m
|
This command sets the gas units to ppb or μg/m
the gas units to μg/m
3
.
Send:
set gas unit ug/m3
Receive:
set gas unit ug/m3 ok
C-Link Protocol Commands
Measurement Configuration
3
). The example
3
. The example below sets
Model 42i Trace Level Instruction Manual
B-33

Advertisement

loading