Choose Signal to Output

Thermo Fisher Scientific
Output
Everything Else
The Choose Signal to Output screen displays a submenu list of the analog
output signal group choices. Group choices are Concentrations, Other
Measurements, and Analog Inputs (if the I/O expansion board option is
installed). This allows the user to select the output signal to the selected
output channel. In dual or auto range mode, "HI" or "LO" is displayed to
indicate high or low concentrations. Range (NO
range mode. The Concentrations screen is shown below. See
below for a list of items for each signal group choice.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > Select Channel > Choose Signal to Output.
CHOOSE SIGNAL TYPE:
>CONCENTRATIONS
OTHER MEASUREMENTS
ANALOG INPUTS
RANGE
AVG
CHOOSE SIGNAL -
SELECTED OUTPUT:
CURRENTLY: NO
SET TO: NONE
CHANGE VALUE
RANGE
AVG
Table 3–7. Signal Type Group Choices
Concentrations
None
NO (single/auto range mode)
NO
(single/auto range mode)
2
NO
(single/auto range mode)
x
LO NO (dual mode)
Zero % Value
0 Units
DIAGS ALARM
CONC
V1
SAVE
DIAGS ALARM
Other Measurements
None
Internal Temp
Chamber Temp
Cooler Temp
NO
Converter Temp
2
Model 42i Trace Level Instruction Manual
Operation
Instrument Controls Menu
Full-Scale 100% Value
10 Units
) is visible only in auto
x
Table 3–7
?
Analog Inputs
None
Analog Input 1
Analog Input 2
Analog Input 3
Analog Input 4
3-49