Servicing

Vacuum Pump Diaphragm and Valve Replacement

Vacuum Pump
Diaphragm and Valve
Replacement
7-10
Model 42i Trace Level Instruction Manual
Figure 7–4. Replacing the Pump
2. Remove both inlet and exhaust lines from the pump.
3. Install the new pump by following the previous steps in reverse.
Use the following procedures to replace the diaphragm and valve for the
vacuum pumps
(Figure
Equipment Required:
Allen wrench, 4 mm
Nut driver, 7/32-inch or 5.5 mm
Allen wrench, 3 mm
Spanner wrench, 3.8 mm diameter by 4.5 mm long inserts
Small flat-blade screwdriver
Large flat-blade screwdriver
Cleaning agent (alcohol)
Fine-grade steel wool
"LOC-TITE" blue breakable thread adhesive
7–5).
Thermo Fisher Scientific