Ozonator Assembly Replacement - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Ozonator Assembly
Replacement
Thermo Fisher Scientific
Use the following procedure to replace the ozonator assembly
13).
Equipment Required:
Ozonator assembly
Wrench, 5/8-inch
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
1. Refer to
Partition
proceed to the next step below.
2. Carefully disconnect the plumbing at the glass inlet and outlet of the
ozonator.
3. Disconnect the stainless steel tubing from the flow transducer.
4. Loosen the four captive screws securing the ozonator bracket to the
floor plate.
Captive Screws (4)
"Removing the Measurement Bench and Lowering the
Panel" in this chapter to lower the partition panel, then
Ozonator – Top View

Ozonator Assembly Replacement

(Figure 7–
Flow Transducer
Model 42i Trace Level Instruction Manual
Servicing
7-25