Table B-10. Analog Voltage Output Range Values - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Thermo Fisher Scientific
set analog vout range channel range
This command sets analog voltage output channel to the range, where
channel is between 1 and 6 inclusive, and range is set according to
10. The example below sets channel 2 to the 0-10 V range.
Send:
set analog vout range 2 3
Receive:
set analog vout range 2 3 ok
Table B–10. Analog Voltage Output Range Values
Range
1
2
3
4
0 [cannot be set to this, but may report]
dig in
This command reports the status of the digital inputs as a 4-digit
hexadecimal string with the most significant bit (MSB) being input 16.
Send:
dig in
Receive:
dig in 0xff7f
din channel
This command reports the action assigned to input channel and the
corresponding active state. The example below reports the input 1 to be
assigned an index number 3 corresponding to action of "set background"
mode with the active state being high.
Send:
din 5
Receive:
din 5 9 AOUTS TO ZERO high
set din channel index state
This command assigns digital input channel (1-16) to activate the action
indicated by index (1-35), when the input transitions to the designated state
(high or low). Use "list din var" command to obtain the list of supported
index values and corresponding actions.
Send:
set din 1 3 high
Receive:
set din 1 3 high ok
C-Link Protocol Commands
Output Range
0-1 V
0-100 mV
0-10 V
0-5 V
Undefined
Model 42i Trace Level Instruction Manual
I/O Configuration
Table B–
B-43