Thermo Scientific 42i Instruction Manual page 259

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Thermo Fisher Scientific
NO
, and prereactor when operating in dual or auto range mode. The
x
example below shows that the NO concentration is 67.2 ppb.
Send:
no
Receive:
no 6720E-2 ppb
conv temp
This command reports the current NO
example below reports that the current converter temperature is 320.7 °C.
Send:
conv temp
Receive:
conv temp 320.7 deg C
cooler temp
This command reports the current PMT cooler temperature. The example
below reports that the current PMT cooler temperature is -2.8 °C.
Send:
pmt temp
Receive:
pmt temp -2.8 deg C
flow
sample flow
These commands report the current measured flow. The example below
reports that the flow measurement is 0.700 liters/minute.
Send:
flow
Receive:
flow 0.7 1/m
internal temp
This command reports the current internal instrument temperature. The
first reading is the temperature being used in instrument calculations. The
second temperature is the actual temperature being measured. If
temperature compensation is on, then both temperature readings are the
same. If temperature compensation is off, a temperature of 30 °C is used as
the default temperature even though the actual internal temperature is 27.2
°C. The example below shows that temperature compensation is on and
that the internal temperature is 27.2 °C.
Send:
internal temp
Receive:
internal temp 027.2 deg C, actual 027.2
pmt temp
This command reports the PMT cooler temperature. The example below
reports that the PMT cooler temperature is -2.8 °C.
Send:
pmt temp
Receive:
pmt temp -2.8 deg C
C-Link Protocol Commands
converter temperature. The
2
Model 42i Trace Level Instruction Manual
Measurements
B-11