Model 42i Trace Level Instruction Manual
In addition to each channel having two ranges, each channel has two span
coefficients. There are two span coefficients so that each range can be
calibrated separately. This is necessary if the two ranges are not close to one
another. For example, the low NO range is set to 0–50 ppb and the high
NO range is set to 0–200 ppb.
By default, in the dual range mode, the analog outputs are arranged on the
rear panel connector as shown in
and pin connections. Dual range mode may be selected from the
Mode
Select" on page 3-74.
Figure 3–5. Pin-Out of Rear Panel Connector in Dual Range Mode
Table 3–3. Default Analog Outputs in Dual Range Mode
Channel
Connector Pin
1
14
2
33
3
15
4
34
5
17
6
36
Ground
16, 18, 19, 35, 37
Note All channels are user definable. If any customization has been made
to the analog output configuration, the default selections my not apply.