Download Print this page

Figure 7-13. Replacing The Ozonator Assembly; Ozonator Transformer Replacement - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Servicing

Ozonator Transformer Replacement

Ozonator Transformer
Replacement
7-26
Model 42i Trace Level Instruction Manual
Ozonator
Ozonator Transformer
Screws (4)
Figure 7–13. Replacing the Ozonator Assembly
5. Remove the two screws securing the ozonator to the ozonator bracket.
6. Unplug the ozonator from the ozonator transformer by lifting the
ozonator straight up.
7. To install the ozonator, follow the previous steps in reverse.
8. Re-install the measurement bench. Refer to
Measurement Bench and Lowering the Partition
Use the following procedure to replace the ozonator transformer
13).
Equipment Required:
Ozonator transformer
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
Ozonator – Side View
Ozonator Bracket Screws (2)
Ozonator Transformer
"Removing the
Panel" in this chapter.
(Figure 7–
Thermo Fisher Scientific

Advertisement

loading