Table B-7. Contrast Levels - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Thermo Fisher Scientific
Send:
contrast
Receive:
contrast 5:50%
set contrast level
This command sets the screen's level of contrast, according to
The example below sets the contrast level to 50%.
Send:
set contrast 5
Receive:
set contrast 5 ok
Table B–7. Contrast Levels
Level
Contrast Level
0
0%
1
10%
2
20%
3
30%
4
40%
5
50%
6
60%
7
70%
8
80%
9
90%
10
100%
conv set temp
This command reports the temperature that the NO
The example below reports that the converter temperature is set to 325 °C.
Send:
conv set temp
Receive:
conv set temp 325 deg C
set conv set temp value
This command sets the temperature that the NO
value is an integer representing dwgrees C. The example below sets the
converter temperature to 325 °C.
Send:
set conv set temp
Receive:
set conv set temp 325 deg C ok
date
This command reports the current date. The example below reports the
date as December 1, 2004.
C-Link Protocol Commands
Hardware Configuration
Table
converter is set to.
2
converter is set to, where
2
Model 42i Trace Level Instruction Manual
B–7.
B-35