Table B-4. Record Output Formats - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

C-Link Protocol Commands
Datalogging
B-22
Model 42i Trace Level Instruction Manual
These commands set the output format for long and short records, and
dynamic data, according to
record output format to ASCII with text.
Send:
set lrec format 1
Receive:
set lrec format 1 ok
Table B–4. Record Output Formats
Format
Output Format
0
ASCII no text
1
ASCII with text
2
Binary data
erec layout
lrec layout
srec layout
These commands report the layout (string indicating the data formats) for
data that is sent out in response to the erec, lrec, srec, and related
commands. For details on how to interpret the strings, see "Record Layout
Definition" later in this appendix.
Send:
lrec layout
Receive:
lrec layout %s %s %lx %f %f %f %f %f %f %f %f %f %f %f
%f
t D L ffffffffffff
flags no nox hino hinox pres pmtt intt rctt convt smplf
ozonf pmtv
lrec mem size
srec mem size
These commands report the long and short records that can be stored with
the current settings and the number of blocks reserved for long and short
records. To calculate the number of short records per block, add 2 to the
number of records, and then divide by the number of blocks. The example
below shows that 7 blocks were reserved for long records and the maximum
number of long records that can be stored in memory is 1426.
Send:
lrec mem size
Receive:
lrec mem size 1426 recs, 7 blocks
Table
B–4. The example below sets the long
Thermo Fisher Scientific