set ozonator safety onoff
These commands set the ozonator safety on or off. The example below sets
the ozonator safety off.
Send:
set ozonator safety off
Receive:
set ozonator safety off ok
ozonator status
This command reports the status of the ozonator and safety. The example
below reports that the ozonator is off.
Send:
ozonator status
Receive:
ozonator status off
pmt status
This command reports the status of the PMT on or off. The example
below reports that the PMT is on.
Send:
pmt status
Receive:
pmt status on
set pmt onoff
These commands set the PMT on or off. The example below turns the
PMT off.
Send:
set pmt off
Receive:
set pmt off ok
save
set save params
This command stores all current parameters in FLASH memory. It is
important that each time instrument parameters are changed, that this
command be sent. If changes are not saved, they will be lost in the event of
a power failure. The example below saves the parameters to FLASH
memory.
Send:
set save params
Receive:
set save params ok
time
This command reports the current time (24-hour time). The example
below reports that the internal time is 2:15:30 pm.
Send:
time
Receive:
time 14:15:30