The dry air capillary along with the pump is used to control flow in the dry
air line.
The processor software tasks are organized into four areas:
Instrument Control
Monitoring Signals
Measurement Calculations
Output Communication
Low-level embedded processors are used to control the various functions on
the boards, such as analog and digital I/O and heater control. These
processors are controlled over a serial interface with a single high-level
processor that also controls the front-panel user interface. The low-level
processors all run a common piece of firmware that is bundled with the
high-level firmware and loaded on power-up if a different version is
detected.
Each board has a specific address that is used to identify to the firmware
what functions are supported on that board. This address is also used for
the communications between the low-level processors and the high-level
processor.
Every tenth of a second the frequency counters, analog I/O, and digital I/O
are read and written to by the low-level processor. The counters are
accumulated over the past second and the analog inputs are averaged over
that second. The high-level processor polls the low-level processors once
per second to exchange the measurement and control data.
Signals are gathered from the low-level processors once per second, and
then processed by the high-level processor to produce the final
measurement values. The one-second accumulated counts representing the
NO/NO
are accumulated and reported for the user-specified averaging
x
time. If this averaging time is greater than ten seconds, the measurement is
reported every 10 seconds. The one-second average of the other analog
inputs are reported directly (no additional signal conditioning is performed
by the high-level processor).
In auto mode, every ten seconds the NO/ NO
processor waits three seconds for the reaction chamber to flush and
solenoid switches and the
x
Thermo Fisher Scientific