Thermo Scientific 42i Instruction Manual page 263

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Thermo Fisher Scientific
Send:
set alarm converter temp max 340
Receive:
set alarm converter temp max 340 ok
alarm cooler temp min
alarm cooler temp max
These commands report the cooler temperature alarm minimum and
maximum value current settings. The example below reports that the cooler
temperature alarm minimum value is -10.0 °C.
Send:
alarm cooler temp min
Receive:
alarm cooler temp min -10.0 deg C
set alarm cooler temp min value
set alarm cooler temp max value
These commands set the cooler temperature alarm minimum and
maximum values to value, where value is a floating-point number
representing cooler temperature alarm limits in degrees C. The example
below sets the cooler temperature alarm maximum value to -2.0 °C.
Send:
set alarm cooler temp max -2.0
Receive:
set alarm cooler temp max -2.0 ok
alarm internal temp min
alarm internal temp max
These commands report the internal temperature alarm minimum and
maximum value current settings. The example below reports that the
internal temperature alarm minimum value is 15.0 °C.
Send:
internal temp alarm min
Receive:
internal temp alarm min 15.0 deg C
set internal temp alarm min value
set internal temp alarm max value
These commands set the internal temperature alarm minimum and
maximum values to value, where value is a floating-point number
representing internal temperature alarm limits in degrees C. The example
below sets the internal temperature alarm maximum value to 45.0 °C.
Send:
set internal temp alarm max 45
Receive:
set internal temp alarm max 45 ok
alarm pressure min
alarm pressure max
These commands report the pressure alarm minimum and maximum value
current settings. The example below reports that the pressure alarm
minimum value is 205 mmHg.
C-Link Protocol Commands
Model 42i Trace Level Instruction Manual
Alarms
B-15