Auto/Manual Mode; Datalogging Settings - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Auto/Manual Mode

Datalogging Settings

Thermo Fisher Scientific
PMT SUPPLY:
CURRENTLY:
SET TO:
RANGE
AVG
The Auto/Manual Mode screen allows selection of the automatic mode
(NO/NO
), NO mode (manual NO), NO
x
prereactor mode (manual prereact). The auto cycle mode switches the
mode solenoid valve automatically on a 10 second cycle so that NO, NO
and NO
concentrations are determined. The manual NO mode puts the
x
mode solenoid valve into the open position so that the sample gas bypasses
the NO
-to-NO converter. Therefore, only the NO concentration is
2
determined. The manual NOx mode puts the mode solenoid valve into the
closed position so that the sample gas passes through the NO
converter. Therefore, only the NO
manual modes, additional averaging times of 1, 2, and 5 seconds are
available from the Averaging Times screen.
In the Main Menu, choose Instrument Controls > Auto/Manual
Mode.
MODE:
CURRENTLY:
SET TO:
RANGE
AVG
The Datalogging Settings menu deals with datalogging.
In the Main Menu, choose Instrument Controls > Datalogging
Settings.
OFF
ON ?
TOGGLE VALUE
DIAGS ALARM
mode (manual NO
x
concentration is determined. In the
x
MANUAL NO
NO/NOX ?
CHANGE VALUE
SAVE VALUE
DIAGS ALARM
Model 42i Trace Level Instruction Manual
Operation
Instrument Controls Menu
) or
x
,
2
-to-NO
2
3-27