Setup Procedure; Figure 2-1. Remove The Packing Material - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Installation

Setup Procedure

Setup Procedure
2-2
Model 42i Trace Level Instruction Manual
Units without Optional I/O Board
Figure 2–1. Remove the Packing Material
4. Check for possible damage during shipment.
5. Check that all connectors and circuit boards are firmly attached.
6. Re-install the cover.
7. Remove any protective plastic material from the case exterior.
8. Remove the external pump from its shipping container and place next
to the instrument.
Use the following procedure to setup the instrument:
1. Connect the sample line to the SAMPLE bulkhead on the rear panel
(Figure
wet, or incompatible materials. All tubing should be constructed of
FEP Teflon®, 316 stainless steel, borosilicate glass, or similar tubing
with an OD of 1/4-inch and a minimum ID of 1/8-inch. The length of
the tubing should be less than 10 feet.
Remove Packing
(2 pieces)
2–2). Ensure that the sample line is not contaminated by dirty,
Units with Optional I/O Board
Remove Packing
(2 pieces)
Thermo Fisher Scientific