Min and Max Pressure
Thermo Fisher Scientific

Pressure

The Pressure screen displays the current reaction chamber pressure reading
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 200 to 450 mmHg. If the pressure reading goes beyond either
the minimum or maximum limit, an alarm is activated. The word
"ALARM" appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Pressure.
The Minimum Pressure alarm limit screen is used to change the minimum
temperature alarm limit. The minimum and maximum pressure screens
Limits
function the same way.
In the Main Menu, choose Alarms > Pressure > Min or Max.
CONVERTER TEMPERATURE:
ACTUAL MIN
SET MIN TO:
INC/DEC
SAVE VALUE
RANGE
AVG
DIAGS ALARM
PRESSURE:
ACTUAL
>MIN
MAX
RANGE
AVG
DIAGS ALARM
PRESSURE:
ACTUAL MIN:
SET MIN TO:
INC/DEC
SAVE VALUE
RANGE
AVG
DIAGS ALARM
o
300.0
C
o
310.0
C ?
232.5 mmHg
200.0 mmHg
450.0 mmHg
200.0 mmHg
210.0 mmHg?
Model 42i Trace Level Instruction Manual
Operation
Alarms Menu
3-67