Other Measurements

Analog Inputs

RS-232/RS-485 Selection
Thermo Fisher Scientific
The Other Measurements screen allows the user to select the output signal
that is tied to the selected streaming data item. The selected item is shown
by "<--" after it. Items displayed are determined by the options installed.
In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > Select Item > Other
Measurements.
OTHER MEASUREMENTS:
>NONE
INT TEMP
CHAMBER TEMP
COOLER TEMP
NO2 CNV TEMP
CHAMBER PRES
FLOW
RANGE
AVG
PMT VOLTS
OZONATOR FLOW
The Analog Inputs screen allows the user to select the analog input signal
(none or analog inputs 1-8) that is tied to the selected streaming data item.
The selected item is shown by "<--" after it.
In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > Select Item > Analog Inputs.
ANALOG INPUTS:
>NONE
ANALOG IN 1
ANALOG IN 2
ANALOG IN 3
ANALOG IN 4
ANALOG IN 5
ANALOG IN 6
RANGE
AVG
The RS-232/RS-485 Selection screen allows the user to choose between the
RS-232 or RS-485 specification for serial communication.
Note Disconnect the serial cable before changing RS-232 and RS-485
selection to prevent damage to the connected equipment.
DIAGS ALARM
DIAGS ALARM
Model 42i Trace Level Instruction Manual
Operation
Instrument Controls Menu
3-39