Download Print this page

Thermo Scientific 42i Instruction Manual page 266

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

C-Link Protocol Commands
Datalogging
B-18
Model 42i Trace Level Instruction Manual
set clr lrecs
set clr srecs
These commands will clear only the long records or only the short records
that have been saved. The example below clears short records.
Send:
set clr srecs
Receive:
set clr srecs ok
data treatment lrec
data treatment srec
These commands report the current selection of data treatment for
concentrations in the long records (lrecs) or short records (srecs). The
example below reports the data treatment for concentrations in lrec is
minimum.
Send:
data treatment lrec
Receive:
data treatment lrec min
set data treatment lrec string
set data treatment srec string
string = | cur | avg | min | max |
These commands set the data treatment to current, average, minimum, or
maximum for the concentration values recorded in the long records (lrecs)
or short records (srecs). The example below sets the data treatment for
concentrations in lrec to minimum.
Send:
set data treatment lrec min
Receive:
set data treatment lrec min ok
list din
list dout
These commands report the current selection for the digital outputs in the
format. Output no Index number variable name active state. The active
state for digital outputs is open or closed. The active state for digital inputs
is high or low.
Send:
list dout
Receive:
list dout
output index variable state
1 35 CONC ALARM open
3 4 UNITS open
4 11 GEN ALARM closed
7 7 NO MODE open
8 8 NOX MODE open
Thermo Fisher Scientific