Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Thermo Fisher Scientific
list lrec
list srec
list stream
list sp
These commands report the list of current selections for long record
logging data, short record logging data, streaming data output, or the
scratch pad (sp) list. The example below shows the list for streaming data
output.
Send:
list stream
Receive:
list stream
field index variable
x x time
1 1 no
2 2 no2
3 3 nox
4 18 intt
5 25 pres
6 26 smplf
er xy
lr xy
sr xy
x = | 0 | 1 |
: Reply termination format (see "set format format"
command)
y = | 0 | 1 | 2 | : Output format (see "set erec/lrec/srec format format"
command)
These commands report the last long and short records stored or the
dynamic data record. In the example below, the command requests a long
record with no checksum, in ASCII format with text. For details on how to
decode the flag fields within these records, see the "flags" command.
Send:
lr01
Receive:
lr01
12:31 02-22-03 flags 54089100 no -8413E-1 nox -8485E-1
lono -6471E-1 lonox -6527E-1 pres 130.9 pmtt 53.2 intt
80.0 rctt 80.0 convt 61 smplf 0.500 ozonf 0.000 pmtv -
115
erec
This command returns a brief description of the main operating conditions
at the time the command is issued (i.e. dynamic data). The example below
shows a typical response. The format is defined by the current settings of
"format" and "erec format" commands. For details on how to decode the
flag fields within these records, see the "flags" command.
Send:
erec
Receive:
erec