This measurement takes several minutes.
10. When the analyzer beeps, press
13. When the single sweep has been taken, press the following keys:
14. Read the mean value and standard deviation from the analyzer display.
Calculate the dynamic range value, using the following equation.
20 x log [(3 x Standard Deviation) + Mean Value]
Write the calculated value on the "Performance 'I&t Record."
15. Repeat the previous two steps (beginning with C-1
marker 1 and marker 2 at the start and stop of the remaining frequency
ranges that are listed on the "Performance 'l&t Record."
proceed as follows:
Repeat steps 6 through 15.
measure port 2.
and repeat steps 6 through 15 to
, Placing